Temperature Calibration
Equipment calibrated: AML Temperature sensors
The calibration uses a GE AS125 high accuracy temperature probe to compare the standard
temperature readings and the raw instrument count which are used to arrive the calibration
coefficients.
Accuracy: Better than +/- 0.003o C
Working Range: -5o C to +40o C